Wash-Sensor WD

Chemical testing system for verifying the performance of washing-disinfectors (WD) in daily routine operations.

  • Easy to use
  • Immediate visual evaluation
  • Complies with legal requirements

Description

With the Wash-Sensor WD, Stericop offers an easy-to-use test for monitoring the cleaning effectiveness of medical instruments in washer-disinfectors (WD). Such a test is required according to RKI/KRINKO 2012* (Appendix 3) for the reprocessing of medical devices.

Our testing system consists of a holder and replaceable indicator plates, which can be reordered separately. The plate features a test soil (red dot) that simulates human blood and tissue residues. In the holder, the test soil is partially covered, mimicking the joint gap of scissors, forceps, or other surgical instruments. This allows you to reliably confirm that the cleaning process has been correctly performed.

Wash-Sensor Holder

You will receive a durable stainless steel holder from us, designed for inserting the indicator plates. Thanks to the practical clip, you can easily attach the test in any WD. Our universal holder is suitable for both Wash-Sensor WD indicator plates and Wash-Sensor U indicator plates.

Product Order Number Pack Size Storage Shelf Life
Wash-Sensor Holder 830001 1 Piece Dry at room temperature 4 years from manufacturing date
Wash-Sensor WD Indicator Plates 831010 10 Pieces
831050 50 Pieces
831100 100 Pieces

Application

The application of the Wash-Sensor WD is quick and straightforward:

  1. Insert the indicator disc sideways into the holder, ensuring that the red test soil remains partially visible.
  2. Place or hang the Wash-Sensor inside the WD, with the open front facing into the WD.
  3. After the cleaning program has completed, remove the Wash Sensor from the WD and pull the indicator disc out of the holder for verification.

Note: For daily routine checks, the testing system should only be used with the normal loading of the WD. Please also observe the manufacturer’s instructions for the correct loading of the WD.

Evaluation

The evaluation of the test is conducted visually by inspecting the indicator strip.

The complete removal of the red dot indicates an effective cleaning process. If the red dot is not completely removed, there may be deficiencies in the mechanical and chemical performance of the WD.

Auswertung des Wash-Sensor RDG

Downloads

Our product flyer contains all the important information about the Wash-Sensor WD, along with a printable illustrated guide.

Purchase the Wash-Sensor WD by Stericop now at our partner shops

    Your search query returned no results.
    If you cannot find your shop, send us a message at info@stericop.com.

    The list could not be loaded.

    WordPress Cookie Plugin by Real Cookie Banner